Counting Chamber Depth and Factor: A Quick Reference Card

A reference card for the bench. It covers the three common depths, the volume each holds and the factor that turns a count into cells per millilitre. Check it against your own validated method before use.
Depth, volume and factor
| Depth | Volume over 1 mm x 1 mm | Multiply the count by | Sample |
|---|---|---|---|
| 100 µm | 100 nl | 10,000 | Diluted and fixed |
| 20 µm | 20 nl | 50,000 | Undiluted |
| 10 µm | 10 nl | 100,000 | Undiluted |
Worked examples
- 100 µm chamber, 200 cells in one grid, sample diluted 1 in 20: 200 x 10,000 x 20 = 40 million per ml.
- 10 µm chamber, 40 cells over the same area, undiluted: 40 x 100,000 = 4 million per ml.
How many to count
- The WHO laboratory manual (sixth edition, 2021) states 200 spermatozoa per replicate, in two replicates.
- Approximate sampling error: 14% at 50 cells, 10% at 100, 7% at 200, 5% at 400.
- Accept the replicates only if they agree within the manual's limit for that total.
Before every count
- Chamber and cover clean and dry.
- Sample well mixed.
- Chamber filled by capillary action, not overfilled.
- Deep chamber left level until the cells have settled.
- Factor on the worksheet matches the chamber on the stage.
The full guide explains how these chambers work and the types available: What Is a Counting Chamber? Depths, Types and How to Choose One.